Endress+Hauser Cerabar PMP55
Absolute and gauge pressure

Endress+Hauser Cerabar PMP55

Pressure transducer

Accuracy

Standard 0.15%
Platinum 0.075%

Process temperature

-70В°...400В°C
(-94В°F...752В°F)

Pressure measuring range

1 bar...400 bar
(15 psi...6000 psi)

Process pressure absolute / max. overpressure limit

600 bar (9000 psi)

Main wetted parts

316L
AlloyC
Montel
Tantal
Rhodium

Max. measurement distance

4000m (13.123ft) H2O

Material process membrane

316L, AlloyC,
Tantal, PTFE,
Rhodium>Gold

Measuring cell

400 mbar...400 bar
(15 psi...6000 psi)
relative/ absolute

Item Number: Cerabar PMP55
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Pressure sensor endress hauser Cerabar M PMP55 has fully welded diaphragm seal for accurate measurements in liquids and gases.

It has found its way into various industries and has proven to be an excellent instrument for measuring mass, level, pressure, volume in liquids.

The device is excellent for high pressure and high process temperature applications: - 70 - + 400°C.

Easy commissioning is ensured by the setup menu, which is provided with an adjustable measuring range. This makes it possible to save time and reduce costs. SIL2 complies with IEC61508 and IEC61511 (IEC61508 and IEC61511).

The main advantages of the pmp55 series:

  • Has a very wide variety of membrane materials and process connections.
  • Availability of a new membrane  TempC makes it possible to minimize the impact of ambient and process temperature changes.
  • Commissioning is quick and easy. Carried out via local menu-driven display, 4 ... 20mA using HART, PROFIBUSPA, FOUNDATION Fieldbus protocols.
  • Easy replacement of the electronic insert and display possible thanks to the modular concept.
  • It has standard ( HART/ PA/ FF) digital protocols for system integration.
  • Single-chamber housing made of stainless steel or aluminum.
  • Hygienic, threaded or flanged process connections available.
  • Measuring range – from -1/0mbar to + 400bar.
  • International certificates of explosion protection, SIL, WHG protection against overflow, hygienic certificates, marine certificates.  

Digital pressure transmitter with fully welded diaphragm seal for measurement in gases or liquids.

Measuring principle

Absolute and gauge pressure

Characteristic / Application

Smart and reliable pressure transmitter, with piezoresistive measuring cell
and metal welded process isolating
diaphragm and diaphragm seal.

Supply / Communication

4 ..20 mA HART:
11,5...45 V DC
Ex ia: 11,5...30 V DC
PROFIBUS PA
FOUNDATION Fieldbus

Accuracy

Standard 0.15%
Platinum 0.075%

Long term stability

< 0,1% von URL/ year
0,2% of URL/ 5 years
0,25% of URL/ 10 years

Ambient temperature

-40В°C...85В°C
(-40В°F...185В°F)

Process temperature

-70В°C...400В°C
(-94В°F...752В°F)

Process pressure absolute / max. overpressure limit

600 bar (9000 psi)

Pressure measuring range

1 bar...400 bar
(15 psi...6000 psi)

Main wetted parts

316L
AlloyC
Montel
Tantal
Rhodium

Process connection

Diaphragm seal
Flanges (DIN, ASME, JIS)
Tri-Clamp ISO2852
Hygienic connections

Max. measurement distance

4000m (13.123ft) H2O

Communication

4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
IO-Link

Certificates / Approvals

ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI

Safety approvals

SIL

Design approvals

EN 10204-3.1
NACE MR0175, MR0103
AD2000

Hygienic approvals

3A, EHEDG
CoC ASME-BPE

Marine approval

GL/ ABS/ LR/ BV/ DNV

Options

local display

Successor

PMP51B

Application limits

Measuring cell: Metal welded
If pressurized, possibly use differential pressure meas-urement with two pressure transmitters (electronic dp). Observe ratio head pressure : hydrostatic pressure

Measuring principle

Absolute and gauge pressure

Characteristic

Smart and reliable pressure transmitter, with piezoresistive measuring cell
and metal welded process isolating diaphragm and diaphragm seal.

Supply voltage

4...20 mA HART
11,5...45V DC (Non Ex):
Ex ia: 11,5...30V DC
PROFIBUS PA:
9...32 V DC (Non Ex)
FOUNDATION Fieldbus:
9...32 V DC (Non Ex)

Reference Accuracy

Standard 0.15%
Platinum 0.075%

Long term stability

0.1% of URL/ year
0.2% of URL/ 5 years
0.25% of URL/ 10 years

Process temperature

-70В°C...+400В°C
(-94В°F...+752В°F)

Ambient temperature

-40В°C...+85В°C
(-40В°F...+185В°F)

Measuring cell

400 mbar...400 bar
(15 psi...6000 psi)
relative/ absolute

Smallest calibratable span

20 mbar (0.3 psi)

Vacuum resistance

50 mbar (0.73 psi)

Max. Turn down

20:1

Max. overpressure limit

600 bar (9000 psi)

Process connection

Thread:
G1/2...G2, MNPT1/2...MNPT2
Flange:
DN25...DN100,
ASME 1"...4",
JIS 10k
Diaphragm seal

Process connection hygienic

Tri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent
SMS
DRD
Universal adapter

Material process membrane

316L, AlloyC,
Tantal, PTFE,
Rhodium>Gold

Material gasket

None, diaphragm welded

Fill fluid

Silicone oil,
Inert oil,
Vegetable oil,
High temperature oil,
Low temperature oil,

Material housing

Die-cast aluminum,
AISI 316L

Communication

4...20 mA
4...20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
IO-Link

Certificates / Approvals

ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI

Safety approvals

SIL

Design approvals

EN10204-3.1
NACE MR0103

Hygienic approvals

CoC ASME-BPE
3A, EHEDG

Marine approvals

GL/ ABS/ LR/ BV/ DNV

Specialities

TempC Membrane

Successor

PMP51B

Comments

No comments yet