Endress+Hauser Cerabar PMP71
Absolute and gauge pressure

Endress+Hauser Cerabar PMP71

Pressure transducer

Accuracy

Standard: 0.05%
Platinum: up to 0.025%

Process temperature

-40В°C...125В°C
(-40В°F...257В°F)

Pressure measuring range

100mbar...700bar
(1.5psi...10.500psi)

Main wetted parts

Alloy C276
316L
Rhodium

Material process membrane

316L, AlloyC,
Rhodium> Gold

Measuring cell

100 mbar...700 bar
(1.5 psi...10.500 psi)
relative/ absolute

Item Number: Cerabar PMP71
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Intelligent gauge and absolute pressure transmitter Cerabar S PMP71 with metal diaphragms for measurements in tanks and pipelines.

Digital pressure transmitter with welded metal sensor for measurement in gases, steam or liquids.

Measuring principle

Absolute and gauge pressure

Characteristic / Application

Digital transmitter with piezoresistive sensor and welded metallic membrane
Modular transmitter
Long term stability
Enhanced safety via self diagnostic functions
Secondary process barrier

Specialities

diagnostic functionalities
different languages in software

Supply / Communication

4...20mA HART:
10,5...45V DC
Ex ia: 10,5...30V DC
PROFIBUS PA /
FOUNDATION Fieldbus:
9...32V DC

Accuracy

Standard: 0.05%
Platinum: up to 0.025%

Long term stability

0,05% of URL/year

Ambient temperature

-50В°C...85В°C
(-58В°F...185В°F)

Process temperature

-40В°C...125В°C
(-40В°F...257В°F)

Process pressure absolute / max. overpressure limit

1050bar (15,750psi)

Pressure measuring range

100mbar...700bar
(1.5psi...10.500psi)

Main wetted parts

Alloy C276
316L
Rhodium

Process connection

Threads
Flanges (DIN, ASME, ...) with flush membrane

Max. measurement distance

7000m (22.966ft) H2O

Communication

4...20 mA HART
1- 5V DC Low Power
PROFIBUS PA
FOUNDATION Fieldbus

Certificates / Approvals

ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC

Safety approvals

SIL

Design approvals

EN 10204-3.1
NACE MR0175, MR0103

Marine approval

GL/ ABS

Drinking water approvals

NSF

Options

HistoROM/M-Dat
4-line digital display
SS- or Aluminiumhousing
Separate housing

Successor

PMP71B

Application limits

Measuring cell: Metal
welded
If pressurized, possibly
use differential pressure
meas-urement with two
pressure transmitters
(electronic dp). Observe
ratio head pressure :
hydrostatic pressure

Measuring principle

Absolute and gauge pressure

Characteristic

Digital transmitter with piezoresistive sensor and welded metallic membrane
Modular transmitter
Long term stability
Enhanced safety via self diagnostic functionsSecondary process barrier

Supply voltage

4...20 mA HART
10,5...45V DC (Non Ex):
Ex ia: 10,5...30V DC
PROFIBUS PA:
9...32 V DC (Non Ex)
FOUNDATION Fieldbus:
9...32 V DC (Non Ex)

Reference Accuracy

Standard: 0.05%
Platinum: up to 0.025%

Long term stability

0.05 % of URL/ year
0.07 % of URL/ 5 years
0.1 % of URL/ 10 years

Process temperature

-40В°C...125В°C
(-40В°F...257В°F)

Ambient temperature

-50В°C...85В°C
(-58В°F...185В°F)

Measuring cell

100 mbar...700 bar
(1.5 psi...10.500 psi)
relative/ absolute

Smallest calibratable span

5 mbar (0.075 psi)

Vacuum resistance

10 mbar (0.15 psi)

Max. Turn down

100:1

Max. overpressure limit

1050 bar (15.750 psi)

Process connection

Thread:
G1/2...G2, R1/2, MNPT1/2...MNPT2, R1/2
Flange:
DN25...DN80,
ASME 1"...4",
JIS 10K...20K

Material process membrane

316L, AlloyC,
Rhodium> Gold

Material gasket

None, measuring cell welded

Fill fluid

Silicone oil,
Inert oil

Material housing

Die-cast aluminum,
AISI 316L

Communication

4...20 mA HART
1- 5V DC Low Power
PROFIBUS PA
FOUNDATION Fieldbus

Certificates / Approvals

ATEX, FM, CSA, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, EAC

Safety approvals

SIL

Design approvals

EN10204-3.1
NACE MR0175

Marine approvals

GL/ABS

Drinking water approvals

NSF

Specialities

Diagnostic functions

Successor

PMP71B

Comments

No comments yet