Endress+Hauser Cleanfit CPA450
Manual retractable assembly

Endress+Hauser Cleanfit CPA450

Retractable assembly for measuring pH and ORP

Process temperature

-15 to 130 В°C (5 to 266 В°F)

Process pressure

Max. 12 bar at 100 В°C (175 psi at 212 В°F)

Item Number: Cleanfit CPA450
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Cleanfit W CPA450 is basic fittings for water and wastewater, chemical, pulp and paper and energy industries.

Basic assembly for the water & wastewater, chemical, pulp & paper and power industries.

Measuring principle

Sensor ORP / Redox

Application

Power stations, chemical industry, industrial water treatment, supply technology

Installation

Manual retractable holder

Characteristic

Open and closed tank, piping

Design

Manual retractable holder for 120mm electrodes, rinse chamber connection G1/4"

Material

Immersion tube: stainless steel 1.4404, Alloy C22, titanium 3.7035
O-rings: EPDM, Viton or Kalrez

Dimension

Immersion depth: up to 700 mm (27.56 inch)

Process temperature

-15 to 130 В°C (5 to 266 В°F)

Process pressure

Max. 12 bar at 100 В°C (175 psi at 212 В°F)

Connection

Thread G 1 1/4", thread NPT 1 1/4"flange DN32 PN16; flange ASME 1 1/4"

Additional certifications

3.1B EN 10204

Measuring principle

Potentiometric

Application

Power stations, chemical industry, industrial water treatment, supply technology

Installation

Manually retractable assembly

Characteristic

Open and closed tank, piping

Design

Manually retractable assembly for 120 mm electrodes, rinse chamber connection G1/4"

Material

Immersion tube: stainless steel 1.4404, Alloy C22, titanium 3.7035
O-rings: EPDM, Viton or Kalrez

Dimension

Immersion depth: up to 700 mm (27.56 inch)

Process temperature

-15 to 130 В°C (5 to 266 В°F)

Process pressure

Max. 12 bar at 100 В°C (175 psi at 212 В°F)

Connection

Thread G 1 1/4", thread NPT 1 1/4"flange DN32 PN16; flange ASME 1 1/4"

Additional certifications

3.1B EN 10204

Comments

No comments yet

Write a comment