Endress+Hauser Condumax CLS16
Analog conductivity sensor

Endress+Hauser Condumax CLS16

Liquid Conductivity Sensor

Measurement range

k=0,1: 0,04ВµS/cm - 500ВµS/cm

Process temperature

Max. 120В°C (150В°C/30 min)
max. 248В°F (302В°F/30 min.)

Process pressure

max. 12 bar at 20В°C
max.174 psi at 68В°F

Item Number: Condumax CLS16
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Condumax H CLS16 is hygienic conductivity probe with fixed cable or interchangeable head. Built-in temperature sensor.

Conductive conductivity sensor for hygienic applications in pure and ultrapure water.

Measuring principle

Conductive

Application

Pharmaceuticals, pure and ultrapure water

Characteristic

Hygienic 2-electrode conductivity sensor

Measurement range

k=0,1: 0,04ВµS/cm - 500ВµS/cm

Measuring principle

Conductive conductivity cell

Design

Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable

Material

Isolation: PEEK
Electrode: polished stainless steel 1.4435
Sealing: Isolast (FFKM) FDA conform

Dimension

Electrode diameter: 17 mm
(0.66 inch)
Electrode length: 54 mm
( 2.10 inch)

Process temperature

Max. 120В°C (150В°C/30 min)
max. 248В°F (302В°F/30 min.)

Process pressure

max. 12 bar at 20В°C
max.174 psi at 68В°F

Temperature sensor

Integrated Pt100 or Pt1000

Ex certification

ATEX

Connection

Process connection: Varivent, Neumo, Clamp
Cable: TOP68 connector or fixed cable

Ingres protection

IP68

Additional certifications

- calibration certification - surface roughness Ra<0,8 (0,4)- Material certification 3.1.B - EHEDG, pharma certificate of compliance

Comments