Endress+Hauser Condumax CLS16D
Digital conductivity sensor
k=0,1: 0,04 to 500 ВµS/cm
Process temperature-5 to 120 В°C (23 to 248 В°F)
For Sterilizsation: max. 150 В°C at 5 bar
(Max. 302 В°F at 73 psi)
13 bar at 20 В°C (188 psi at 68 В°F)
9 bar at 120 В°C (130 psi at 248 В°F)
- Overview
- Comments
Memosens conductive conductivity sensor for hygienic applications in pure and ultrapure water.
Measuring principleConductive
ApplicationPure and ultrapure water, pharmaceutical industry
CharacteristicHygienic 2-electrode conductivity sensor
Measurement rangek=0,1: 0,04 to 500 ВµS/cm
Measuring principleConductivity cell
DesignHygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable
MaterialIsolation: PEEK
Electrode: polished stainless steel 1.4435
Sealing: Isolast (FFKM) FDA conform
Electrode diameter: 17 mm (0.66 inch)
Electrode length: 54 mm (2.13 inch)
-5 to 120 В°C (23 to 248 В°F)
For Sterilizsation: max. 150 В°C at 5 bar
(Max. 302 В°F at 73 psi)
13 bar at 20 В°C (188 psi at 68 В°F)
9 bar at 120 В°C (130 psi at 248 В°F)
NTC 30K
ConnectionProcess connection: Varivent, Neumo, Clamp
Cable: Memosens connector
IP68
Additional certificationsMaterial certification 3.1
EHEDG certified, hygienic design, CoC ASME BPE
Conformity to cGMP derived requirements
ASME BPE
FDA approved
Regulation (EC) No.
1935/2004
China Food
Comments
No comments yet