Endress+Hauser Memosens COS22E
Digital oxygen sensor
Standard sensor: 0 to 60 mg/l, 0 to 600 %SAT, 0 to 1200 hPa, 0 to 100 Vol%
Trace sensor: 0 to 10 mg/l, 0 to 120 %SAT, 0 to 250 hPa, 0 to 25 Vol%
-5 to 135 В°C (23 to 275 В°F)
Process pressure1 to 12 bar abs (14.5 to 174 psi)
- Overview
- Comments
Memosens 2.0 hygienic, amperometric oxygen sensor for the food and life sciences industries.
Measuring principleAmperometric oxygen measurement
ApplicationTypical applications are:
- Pharmaceuticals and biotechnology: process control in enzyme production, control of culture preparations
.
- Water treatment: boiler feed water, WFI (Water for Injection)
- Residual oxygen measurement in processes
Installation
CharacteristicDigital hygienic amperometric oxygen sensor
with maximum measurement stability over several
sterilization cycles available in two measuring ranges
Standard sensor: 0 to 60 mg/l, 0 to 600 %SAT, 0 to 1200 hPa, 0 to 100 Vol%
Trace sensor: 0 to 10 mg/l, 0 to 120 %SAT, 0 to 250 hPa, 0 to 25 Vol%
Measuring principle
DesignDesign
MaterialSensor shaft: Stainless steel 1.4435 (AISI 316L)
Process seal: FKM (USP<87>, <88> Class VI and FDA)
Process seal for Ex versions: FKM (not FDA compliant)
Seals/O-rings: EPDM (USP<87>, <88> Class VI and FDA), FKM
(FDA)
Shaft sleeve: Stainless steel 1.4435 (AISI 316L) or
Titanium or Hastelloy
Membrane cover layer: Silicone (USP<87>, <88> Class VI and FDA)
Diameter: 12 mm (0.47 inch)
Shaft length: 120, 160, 220, 320, 420 mm
(4.72, 6.30, 8.66, 12.60 und 16.54 inch)
-5 to 135 В°C (23 to 275 В°F)
Process pressure1 to 12 bar abs (14.5 to 174 psi)
Temperature sensorNTC 22k
ConnectionInductive, contactless connection head with Memosens 2.0 technology
Ingres protectionIP68
Additional certificationsAdditional certifications
Comments
No comments yet