Endress+Hauser Liquiline System CA80FE
Iron analyzer
0.05 to 2 mg/l Fe
0.1 to 5 mg/l Fe
0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe
4 to 40 В°C (39.2 to 104 В°F)
Process pressureAt atmospheric pressure, < 0.2 bar
- Overview +
- Comments +
Colorimetric system for monitoring of drinking water, mineral water and process water.
Measuring principleColorimetric
CharacteristicProcess analyzer for iron in aqueous solutions
Comply with standard colorimetric measuring principle, ferrozine method
Housing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in
Open design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel
4 to 40 В°C (39.2 to 104 В°F)
Ambient temperature5 to 40 В°C (41 to 104 В°F)
Outdoor version: - 20 to 40 В°C (-4 to 104 В°F)
At atmospheric pressure, < 0.2 bar
Sample flow rateMin. 5 ml/min (0.17 fl.oz/min)
Consistency of the sampleSuspended solids content
Turbitity < 50 NTU, aqueous, homogenized
Monitoring of the iron content of potable water, mineral water and source water for industrial processes
Power supply100 to 120 VAC / 200 to 240 VAC В± 10%
24 VDC В± 10%
50 В± 1 or 60 В± 1,2 Hz
2x 0/4 to 20 mA
Webserver, Modbus, Ethernet/IP, Profibus DP
1 or 2 measuring channel
1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
0.05 to 2 mg/l Fe
0.1 to 5 mg/l Fe
0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe
Reagents and standard solutions CY80FE for the operation
Regular maintenance is done with the parts of the maintenance kit CAV800
Buy Endress+Hauser Liquiline System CA80FE with the best possible price and delivery terms on eltra-trade.com
Find the industrial control components you need at eltra-trade.com. Our catalog includes a variety of sizes and types, all available for fast delivery. Enjoy competitive pricing and excellent customer service. Trust us to fulfill your automation equipment requirements.
Comments