Endress+Hauser CA78
Low-range TOC analyzer

Endress+Hauser CA78
Measurement range

0.5 to 1 000 Ојg/l (ppb)

Process temperature

< 50 В°C (122 В°F)

Process pressure

max. 0.5 bar (7.25 psi)

Measuring method

TOC determination by UV digestion and measurement of the differential conductivity

Item Number: CA78
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Precise online TOC monitoring in power plants and semiconductor production.

Measuring principle

Differential conductivity

Characteristic

Total carbon (TOC) analyzer for trace levels

Measuring method

TOC determination by UV digestion and measurement of the differential conductivity

Size

Housing:
500 x 290 x 200 mm
19.68 x 11.41 x 7.87 in

Design

Stainless steel housing;
IP 42 (standard), IP54 (optional)

Process temperature

< 50 В°C (122 В°F)

Ambient temperature

5 to 50 В°C (23 to 122 В°F)

Process pressure

max. 0.5 bar (7.25 psi)

Sample flow rate

Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the sample

max. conductivity 2 ОјS/cm, optional: max. 10 ОјS/cm; particle free

Specials

UV reactor with continuous function monitoring

Application

Determination of total carbon in ultrapure water applications, e.g. in the power or semiconductor industry, that meet the following conditions:
Conductivity < 10 ОјS/cm
pH range: neutral

Power supply

100/240 V AC, 47 - 63 Hz

Output / communication

0/4 to 20 mA, galvanically isolated

Input

1 to optional 3 measuring channels
Optinonal control input 24 V (for 1 channel instruments)

Measurement range

0.5 to 1 000 Ојg/l (ppb)

Comments

No comments yet

Write a comment