Endress+Hauser Liquiline System CA80PH
Orthophosphate analyzer

Endress+Hauser Liquiline System CA80PH
We ship to United States of America!
Measurement range

0 to 2.5 mg/l PO4-P (blue method)
0.05 to 10 mg/l PO4-P (blue method)
0.05 to 10 mg/l with dilution function to maximum 2.5 to 500 mg/l PO4-P (blue method)
0.5 to 20 mg/l PO4-P (yellow method)
0.5 to 50 mg/l PO4-P (yellow method)
0.5 to 50 mg/l with dilution function to maximum 10 to 1000 mg/l PO4-P (yellow method)

Process temperature

4 to 40 В°C (39.2 to 104 В°F)

Process pressure

At atmospheric pressure, < 0.2 bar

Item Number: Liquiline System CA80PH
Price on Request
-
+
Total price: 0.00
Delivery standard: 3-6 weeks, depending on the country.
Condition: new
Contact our experts
  • Overview +
  • Comments +

Colorimetric system for wastewater monitoring and optimization of process control.

Measuring principle

Colorimetric

Characteristic

Process analyzer for orthophosphate in aqueous solutions
Comply with standard colorimetric measuring principle, molybdenum blue method, following DIN EN 1189 or molybdate vanadate method

Size

Housing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in

Design

Open design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel

Process temperature

4 to 40 В°C (39.2 to 104 В°F)

Ambient temperature

5 to 40 В°C (41 to 104 В°F)
Outdoor version: - 20 to 40 В°C (-4 to 104 В°F)

Process pressure

At atmospheric pressure, < 0.2 bar

Sample flow rate

Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the sample

Suspended solids content
Turbitity < 50 NTU, aqueous, homogenized

Application

Monitoring and optimization of the cleaning capacity of municipal and industrial waste water treatment plants

Power supply

100 to 120 VAC / 200 to 240 VAC В± 10%
24 VDC В± 10%
50 В± 1 or 60 В± 1,2 Hz

Output / communication

2x 0/4 to 20 mA
Modbus RS485/TCP

Input

1 or 2 measuring channel
1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

Measurement range

0 to 2.5 mg/l PO4-P (blue method)
0.05 to 10 mg/l PO4-P (blue method)
0.05 to 10 mg/l with dilution function to maximum 2.5 to 500 mg/l PO4-P (blue method)
0.5 to 20 mg/l PO4-P (yellow method)
0.5 to 50 mg/l PO4-P (yellow method)
0.5 to 50 mg/l with dilution function to maximum 10 to 1000 mg/l PO4-P (yellow method)

Consumables

Reagents and standard solutions CY80H as well as cleaner CY800 (blue method) are necessary for the operation
Regular maintenance is done with the parts of the maintenance kit CAV800


Buy Endress+Hauser Liquiline System CA80PH with the best possible price and delivery terms on eltra-trade.com

Eltra Trade s.r.o. is a global leader in industrial automation supply. Our extensive catalog includes top brands, with worldwide delivery and competitive pricing. We focus on customer satisfaction and reliability. Experience our exceptional service and support for your automation needs.

Comments

No comments yet

Write a comment