Endress+Hauser Proline Promag H 100
Electromagnetic flowmeter

Endress+Hauser Proline Promag H 100
Max. measurement error

Volume flow (standard): В±0.5 % o.r. В± 1 mm/s (0.04 in/s)
Volume flow (option) В±0.2 % o.r. В± 2 mm/s (0.08 in/s)

Measuring range

0.06 dmВі/min to 600 mВі/h (0.015 to 2650 gal/min)

Medium temperature range

20 to +150 В°C (4 to +302 В°F)

Max. process pressure

PN 40, Class 150, 20K

Wetted materials

Liner: PFA
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve
Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone)
Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum

Item Number: Proline Promag H 100
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

The specialist for hygienic applications with an ultra-compact transmitter.

Measuring principle

Electromagnetic

Product headline

Specialist for hygienic applications with an ultra-compact transmitter.
Multivariable measurement of flow, temperature and conductivity.
Dedicated to demanding applications in the food and beverage as well as in life sciences industries.

Sensor features

Flexible installation concept numerous hygienic process connections. Energy...saving flow measurement no pressure loss due to cross section constriction. Maintenance...free no moving parts.
Integrated temperature measurement. Sensor housing made of stainless steel (3-A, EHEDG). Wetted materials CIP, SIP cleanable.

Transmitter features

Space...saving transmitter full functionality on the smallest footprint. Time...saving local operation without additional software and hardware integrated web server. Integrated verification Heartbeat Technology.
Robust, ultra-compact transmitter housing. High degree of protection: IP69. Local display available.

Nominal diameter range

DN 2 to 150 (В№вЃ„в‚Ѓв‚‚ to 6")

Wetted materials

Liner: PFA
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve
Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone)
Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum

Measured variables

Volume flow, temperature, conductivity, mass flow, corrected volume flow, corrected
conductivity

Max. measurement error

Volume flow (standard): В±0.5 % o.r. В± 1 mm/s (0.04 in/s)
Volume flow (option) В±0.2 % o.r. В± 2 mm/s (0.08 in/s)

Measuring range

0.06 dmВі/min to 600 mВі/h (0.015 to 2650 gal/min)

Max. process pressure

PN 40, Class 150, 20K

Medium temperature range

20 to +150 В°C (4 to +302 В°F)

Ambient temperature range

40 to +60 В°C (40 to +140 В°F)

Sensor housing material

1.4301 (304), corrosion resistant

Transmitter housing material

Compact: AlSi10Mg, coated
Compact/ultra...compact: 1.4301 (304)

Degree of protection

Standard: IP66/67, type 4X enclosure
Option: IP69

Display/Operation

4...line backlit display available (no local operation)
Configuration via web browser and operating tools possible

Outputs

4...20 mA HART (active)
Pulse/frequency/switch output (passive)

Inputs

None

Digital communication

HART, PROFIBUS DP, Modbus RS485, EtherNet/IP, PROFINET

Power supply

DC 20 to 30 V

Hazardous area approvals

ATEX, IECEx, cCSAus, INMETRO, EAC

Product safety

CE, C-tick

Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Verification: Heartbeat Technology complies with requirements for traceable verification according to ISO 9001:2008, chapter 7.6. a (TUV attestation)

Marine approvals and certificates

LR approval, DNV GL approval, ABS approval, BV approval

Pressure approvals and certificates

PED, CRN

Material certificates

3.1 material

Hygienic approvals and certificates

3-A, liner and seals acc. to FDA, cGMP

Comments

No comments yet