Endress+Hauser Proline Prosonic Flow E 100
Ultrasonic flowmeter

Endress+Hauser Proline Prosonic Flow E 100
Max. measurement error

Volume flow:
- В± 0.5 % o.r. В± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- В± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)

Measuring range

0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min)

Medium temperature range

0 to 150В° (17.8 to 302 В°F)

Max. process pressure

PN 25 / ASME Cl. 150

Wetted materials

Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105

Item Number: Proline Prosonic Flow E 100
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Economical Ultrasonic flowmeter with integrated temperature measurement.

Measuring principle

Ultrasonic flow

Product headline

Economical Ultrasonic flowmeter with integrated temperature measurement.
Bidirectional measuring of demineralized water applications for Utilities, e.g. boiler condensate return lines.

Sensor features

Long-term stability reliable, robust sensor. Reducing further measuring point multivariable device. Dependable flow measurement high turndown (200:1).
Measurement accuracy up to В±0.5 % (flow) or according to EN 1434 Cl. 2, В±2.0 В°C (В± 3.6 В°F) (temperature). Process temperatures up to 150 В°C (302 В°F). Entire meter housing made of stainless steel.

Transmitter features

Time-saving local operation without additional software and hardware integrated web server. Extended calibration intervals integrated device verification due to Heartbeat Technology. Easy commissioning brief parameter explanations.
4-20 mA HART, pulse/frequency output. Local display for reading and monitoring available. Robust transmitter housing.

Nominal diameter range

DN 50 to 150 (2"to 6")

Wetted materials

Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105

Measured variables

Volume flow, calculated mass flow, flow velocity, sound velocity, temperature

Max. measurement error

Volume flow:
- В± 0.5 % o.r. В± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- В± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)

Measuring range

0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min)

Max. process pressure

PN 25 / ASME Cl. 150

Medium temperature range

0 to 150В° (17.8 to 302 В°F)

Ambient temperature range

-20 to 60 В°C (-4 to 140 В°F)

Sensor housing material

Stainless steel

Transmitter housing material

Compact:
AlSi10Mg, coated
Stainless steel, 1.4301

Degree of protection

Standard: IP66/67, Type 4X enclosure
With opened housing: IP20, Type 1 enclosure

Display/Operation

4...line backlit display available (no local operation)
Configuration via web browser and operating tools possible

Outputs

4...20 mA HART (active)
Pulse/frequency/switch output (passive)

Inputs

None

Power supply

DC 19.2 to 28.8 V

Hazardous area approvals

None

Product safety

CE, C-Tick, EAC marking

Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 Section 7.1.5.2 a (TГњV SГњD attestation)
Calibration according to EN 1434 Class 2

Pressure approvals and certificates

PED

Material certificates

3.1 material

Comments

No comments yet

Write a comment