Endress+Hauser Proline Promag 10D
Electromagnetic flowmeter

Endress+Hauser Proline Promag 10D

Electromagnetic flowmeter

Max. measurement error

Volume flow: В±0.5 % o.r. В± 2 mm/s (В±0.5 % o.r. В± 0.08 in/s)

Measuring range

9 to 4700 dm3/min (2.5 to 1250 gal/min)

Medium temperature range

0 to +60 В°C (+32 to +140 В°F)

Max. process pressure

PN 16, Class 150, 10K

Item Number: Proline Promag 10D
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

Electromagnetic flow meter Promag D. For various tasks of measuring water flow. Lightweight and compact.

The highly cost-effective flowmeter, available as compact wafer.

Measuring principle

Electromagnetic

Product headline

The highly cost-effective flowmeter designed as compact wafer version.
For basic water applications, optimized for limited space and plastic pipe installations.

Sensor features

Easy, fast centering of the sensor innovative housing construction. Energy...saving flow measurement no pressure loss due to cross section constriction. Maintenance...free no moving parts.
Short face-to-face length and low weight. Integrated ground disks made of stainless steel. International drinking water approvals.

Transmitter features

Cost-effective designed for easy applications and direct integration. Safe operation display provides easily readable process information. Fully industry-compliant IEC/EN/NAMUR.
2-line display with push buttons. Device as compact or remote version. HART.

Nominal diameter range

DN 25 to 100 (1 to 4")

Wetted materials

Liner: Polyamide
Electrodes: 1.4435 (316L)

Measured variables

Volume flow

Max. measurement error

Volume flow: В±0.5 % o.r. В± 2 mm/s (В±0.5 % o.r. В± 0.08 in/s)

Measuring range

9 to 4700 dm3/min (2.5 to 1250 gal/min)

Max. process pressure

PN 16, Class 150, 10K

Medium temperature range

0 to +60 В°C (+32 to +140 В°F)

Ambient temperature range

20 to +60 В°C (4 to +140 В°F)

Sensor housing material

AlSi10Mg, coated
Sensor connection housing: AlSi10Mg, coated

Transmitter housing material

Powder...coated die...cast aluminum

Degree of protection

Compact version: IP66&67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP 67, type 4X enclosure

Display/Operation

2...line display with push buttons
Configuration via local display and operating tools possible

Outputs

4...20 mA HART (active)
Pulse/switch output (passive)

Inputs

None

Digital communication

HART

Power supply

DC 11 to 40 V
AC 85 to 250 V (45 to 65 Hz)
AC 20 to 28 V (45 to 65 Hz)

Hazardous area approvals

FM
CSA

Product safety

CE, C-tick, EAC marking

Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)

Hygienic approvals and certificates

Drinking water approval: ACS, KTW/W270, NSF 61, WRAS BS 6920

Comments

No comments yet