Endress+Hauser Micropilot FMR62
Radar measurement

Endress+Hauser Micropilot FMR62
Accuracy

+/- 1 mm (0.04 in)

Process temperature

-40...+200 В°C
(-40...+392 В°F)

Process pressure absolute / max. overpressure limit

Vacuum...+25 bar
(Vacuum...+362.6 psi)

Max. measurement distance

80 m (262 ft)

Main wetted parts

PTFE or PEEK antenna,
316L threads,
PTFE cladded flanges

Item Number: Micropilot FMR62
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

For 80GHz level measurement in aggressive liquids or applications with hygiene requirements.

Measuring principle

Level radar

Characteristic / Application

Premium device for continuous non-contact level measurement,
in which aggressive media are used as well as for highest hygiene requirements (ASME
BPE, USP Class VI); For applications with many obstacles for small vessels due to the very small beam angle.

Specialities

Heartbeat Technology
SIL 2/3 according to IEC 61508,
BluetoothВ® commissioning,
Operation and maintenance SmartBlue App,
Safety and reliability with Multi-Echo Tracking,
HistoROM,
RFID TAG for easy identification

Supply / Communication

2-wire (HART),
BluetoothВ® wireless technology and App (optional)

Frequency

W-band (~80 GHz)

Accuracy

+/- 1 mm (0.04 in)

Ambient temperature

-40...+80 В°C
(-40...+176 В°F)

Process temperature

-40...+200 В°C
(-40...+392 В°F)

Process pressure absolute / max. overpressure limit

Vacuum...+25 bar
(Vacuum...+362.6 psi)

Main wetted parts

PTFE or PEEK antenna,
316L threads,
PTFE cladded flanges

Process connection

Threads:
G, MNPT 3/4", 1-1/2"
Flanges:
DN50 ... DN150
ASME 2" ...6"

Process connection hygienic

DIN11851, Tri-Clamp

Max. measurement distance

80 m (262 ft)

Communication

4...20 mA HART,
Additional switch,
BluetoothВ® wireless technology

Certificates / Approvals

ATEX, FM, CSA C/US, IEC Ex, JPN Ex, INMETRO, NEPSI, KC, EAC, UK Ex

Safety approvals

Overfill protection WHG,
SIL 2, SIL 3

Design approvals

EN 10204-3.1
NACE MR0175, MR0103
AD2000

Hygienic approvals

CoC-ASME BPE

Options

Display,
Customized parameterization,
Remote operation via SmartBlue App using BluetoothВ®,
PWIS free,
Gastight feed through

Application limits

Process temp. > 200 В°C (392 В°F) -> FMR5x
Strong turbulences and foam, stilling well and bypass -> FMR5x,FMP5x

Comments

No comments yet

Write a comment