Endress+Hauser Raman Rxn-41 probe
532 nm, 785 nm, 1000 nm
316L option
Metal: 316L stainless steel
Window: High-purity sapphire
C276 option
Metal: C276 alloy
Window: High-purity sapphire
Titanium option
Metal: Grade 2 titanium
Window: High-purity sapphire
Hybrid metal combination option
Metal: 316L stainless steel, C276 alloy
Window: High-purity sapphire
ATEX, CSA, IECEx
- Overview
- Comments
Offering simplified, reduced cost installation in the process environment.
Measuring principleRaman spectroscopy
Laser wavelength
532 nm, 785 nm, 1000 nm
Probe spectral coverage is limited by the coverage of the analyzer being used.
Maximum laser power into probehead (mW)<499
Sample interfaceTemperature 1" standard configuration:
-30 to 120 В°C (316L stainless steel)
-30 to 150 В°C (C276 alloy)
-30 to 150 В°C (Grade 2 titanium)
Temperature 1" cryogenic configuration:
-196 to 70В°C (C276 alloy)
-196 to 70 В°C (Hybrid metal combination)
Temperature 2.375" standard configuration:
-30 to 120В°C (316L stainless steel)
-30 to 150 В°C (C276 alloy)
Temperature ramp: ≤30 °C/min
Flange: up to 305 mm diameter, ASME B16.5 or DIN EN1092 Type B flanges available upon request
Relative humidity: up to 95%, non-condensing
Min pressure (Bara):
Full vacuum (0 Bara) but not ultrahigh vacuum (UHV) where outgassing maybe a concern
Max Pressure (Barg)
1" probe:
77.0 (316L stainless steel)
98.5 (C276 alloy)
76.0 (Grade 2 titanium)
1" cryogenic probe:
111.0 (C276 alloy)
87.5 (Hybrid metal combination)
2.375" probe (Barg):
42.7 (316L stainless steel)
66.1 (C276 alloy)
316L option
Metal: 316L stainless steel
Window: High-purity sapphire
C276 option
Metal: C276 alloy
Window: High-purity sapphire
Titanium option
Metal: Grade 2 titanium
Window: High-purity sapphire
Hybrid metal combination option
Metal: 316L stainless steel, C276 alloy
Window: High-purity sapphire
Cable sold separately
LengthMaximum immersible length is dependent on material choice
Diameter (mm)25.4 or 60.325
(25.4 only for Grade 2 Titanium)
ATEX, CSA, IECEx
Comments
No comments yet