Endress+Hauser Memosens Wave CAS80E
Spectrometer

Endress+Hauser Memosens Wave CAS80E
Measurement range

"Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l
CODeq: 0 to 20 000 mg/l
BODeq: 0 to 5000 mg/l
SAC254: 0 to 1000 /m
TU: 0 to 800 FAU
TSS: 0 to 10 000 mg/l
NO3-N: 0 to 500 mg/l
APHA/Hazen: 0 to 500 Hazen

Process temperature

0 to 50В°C
32 to 122 В°F

Process pressure

0.5 to 10 bar abs.
7.3 to 145 psi

Measuring method

UV/VIS Absorbance at 200 to 800 nm

Item Number: Memosens Wave CAS80E
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments

UV-VIS spectrometer for the water & wastewater industry.

Measuring principle

Absorption

Application

Monitoring of WWTP in- and outlet
Drinking water and surface water

Installation

Insitu (immersion in process), extractive with sample and flow cell

Characteristic

Insitu (immersion in process), extractive with sample and flow cell
UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color

Measurement range

"Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l
CODeq: 0 to 20 000 mg/l
BODeq: 0 to 5000 mg/l
SAC254: 0 to 1000 /m
TU: 0 to 800 FAU
TSS: 0 to 10 000 mg/l
NO3-N: 0 to 500 mg/l
APHA/Hazen: 0 to 500 Hazen

Measuring principle

UV/VIS absorbance at 200 to 800 nm

Design

40 mm

Material

Housing: stainless steel or titanium
Windows: sapphire or quartz
O-rings: EPDM

Dimension

571.4 x 40 mm
21.5 x 1.6 in

Process temperature

0 to 50В°C
32 to 122 В°F

Process pressure

0.5 to 10 bar abs.
7.3 to 145 psi

Temperature sensor

-20 to 60В°C
-4 to 140 В°F

Connection

G1 and NPT 3/4"

Ingres protection

IP68

Measuring principle

UV absorbance

Characteristic

Insitu (immersion in process), extractive with sample and flow cell UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color

Measuring method

UV/VIS Absorbance at 200 to 800 nm

Size

571.4 x 40 mm
21.5 x 1.6 in

Design

40 mm

Process temperature

0 to 50В°C
32 to 122 В°F

Ambient temperature

-20 to 60В°C
-4 to 140 В°F

Process pressure

0.5 to 10 bar abs.
7.3 to 145 psi

Specials

G1, NPT3/4 IP68 (1m/3,3ft) water column with 1N KCl during 60 days

Application

Monitoring of WWTP in- and outlet
Drinking water and surface water

Measurement range

Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l
CODeq: 0 to 20 000 mg/l
BODeq: 0 to 5000 mg/l
SAC254: 0 to 1000 /m
TU: 0 to 800 FAU
TSS: 0 to 10 000 mg/l
NO3-N: 0 to 500 mg/l
APHA/Hazen: 0 to 500 Hazen

Measuring principle

UV absorbance

Application

Monitoring of WWTP in- and outlet, drinking water and surface water

Characteristic

Insitu (immersion in process)
Extractive with sample and flow cell UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color

Measurement range

Depending on the path length and the application the meausring range varies between: TOCeq: 0 to 400 mg/l
CODeq: 0 to 20 000 mg/l
BODeq: 0 to 5000 mg/l
SAC254: 0 to 1000 /m
TU: 0 to 800 FAU
TSS: 0 to 10 000 mg/l
NO3-N: 0 to 500 mg/l
APHA/Hazen: 0 to 500 Hazen

Design

40 mm

Material

Housing: stainless steel or titanium
Windows: sapphire or quartz
O-rings: EPDM

Process temperature

0 to 50В°C
32 to 122 В°F

Process pressure

0.5 to 10 bar abs.
7.3 to 145 psi

Temperature sensor

-20 to 60В°C
-4 to 140 В°F

Connection

G1 and NPT 3/4"

Ingres protection

IP68

Comments

No comments yet

Write a comment