Endress+Hauser SS2100
Gas analyzer
Concentration
Cell Pressure
Cell Temperature
CSA Class I, Division 2
CSA Class I, Zone 2
- Overview +
- Comments +
Exceptionally reliable for measuring trace gas components.
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for hydrogen sulfide (H2S) measurement. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesH2S (Hydrogen Sulfide): 0-10 to 0-1000 ppmv; 0-5000 ppmv to 0-5%
Measured VariablesConcentration
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for moisture (H2O) measurement. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesH2O (Moisture): 0-10 to 0-100 ppmv; 0-50 to 0-5000 ppmv
Measured VariablesConcentration
Moisture Dew Point
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for carbon dioxide (CO2) measurement. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesCO2 (Carbon Dioxide): 0-10 to 0-1000 ppmv; 0-5000 ppmv to 0-5%
Measured VariablesConcentration
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for ammonia (NH3) measurement. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesNH3 (Ammonia): 0-5 ppmv
Measured VariablesConcentration
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAn advanced gas analyzer for acetylene (C2H2) measurement. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels1
Analyte and Measurement rangesC2H2 (Acetylene): 0-5; 0-3000 ppmv
Measured VariablesConcentration
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs: 2, Flow Alarm, Validation Request
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAdvanced multi-channel gas analyzer for hydrogen sulfide (H2S) and moisture (H2O), also known as a 2-Pack. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels2
Analyte and Measurement rangesH2S (Hydrogen Sulfide): 0-10 to 0-1000 ppmv
H2O (Moisture): 0-100 to 0-5000 ppmv
Concentration
Moisture Dew Point
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Four 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O)
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs (H2S): 2, Flow Alarm, Validation Request
Digital Outputs (H2O): 2, Concentration Alarm, General Fault
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAdvanced multi-channel gas analyzer for hydrogen sulfide (H2S) and carbon dioxide (CO2), also known as a 2-Pack. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels2
Analyte and Measurement rangesH2S (Hydrogen Sulfide): 0-10 to 0-1000 ppmv
CO2 (Carbon Dioxide): 0-5000 ppmv to 0-5%
Concentration
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Four 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (CO2)
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs (H2S): 2, Flow Alarm, Validation Request
Digital Outputs (CO2): 2, Concentration Alarm, General Fault
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Measuring principleTDLAS
Product HeadlineAdvanced multi-channel gas analyzer for hydrogen sulfide (H2S), moisture (H2O), and carbon dioxide (CO2), also known as a 3-Pack. The SS2100 uses Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.
Channels3
Analyte and Measurement rangesH2S (Hydrogen Sulfide): 0-10 to 0-1000 ppmv
H2O (Moisture): 0-100 to 0-5000 ppmv
CO2 (Carbon Dioxide): 0-5000 ppmv to 0-5%
Concentration
Moisture Dew Point
Cell Pressure
Cell Temperature
-20 to 50В°C (-4 to 122В°F)
Optional: -10 to 60В°C (14 to 140В°F)
Inlet Pressure: 140-350 kPa (20-50 psig)
Sample Cell: 800-1200 mbara or 950-1700 mbara (optional)
316L Stainless Steel
FKM O-Rings
Glass
120 or 240 VAC В±10%, 50-60 Hz, 300W
(18-24 VDC - optional for electronics only)
3 Amps max @ 120 VAC , 1.5 Amps max @ 240 VAC Hz
Analog Output: Four 4-20mA Isolated, 1200 ohms @ 24 VDC max load
Serial: RS232C and Ethernet (H2S) and RS232C or Ethernet (H2O/CO2)
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Digital Outputs (H2S): 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
Digital Inputs (H2S): 2, Flow Alarm, Validation Request
Digital Outputs (H2O/CO2): 2, Concentration Alarm, General Fault
Electronics: 304 or 316L Stainless Steel
Sample System Enclosure: 304 or 316L Stainless Steel
CSA Class I, Division 2
CSA Class I, Zone 2
IP66, Type 4X
Buy Endress+Hauser SS2100 with the best possible price and delivery terms on eltra-trade.com