Pepperl+Fuchs LGC
Level probe

Pepperl+Fuchs LGC
  • Hydrostatic pressure sensor for level measuring of water
  • Measuring ranges: 0 bar ... 0.1 bar to 0 bar ... 20 bar
  • High-precision and long-term stability ceramic measuring cell
  • High mechanical resistance to overload and aggressive media
  • Permanent hermetically sealed level probe
  • Electronics comprising 4 mA ... 20 mA output signal and integrated overvoltage protection
  • Simultaneous level and temperature measuring by optional integrated temperature probe Pt100
  • KTW and NSF drinking water approval
Item Number: LGC
Price on Request
-
+
Total price: 0.00
Contact our experts
  • Overview
  • Comments
Measurement rangeLGC:
- nine fixed pressure measuring ranges in bar, see ordering information
- customer-specific measuring ranges, factory-calibrated
Pt100 (optional): temperature measurement from -10 ... 70 В°C (263 ... 343 K)
Adjustment timeLGC: 150 ms, Pt100 (optional): 300 s
Degree of protectionIEC 60529:2001
Output signalLGC: 4 ... 20 mA for hydrostatic pressure measured value, two-wire
Pt100 (optional): temperature-dependent resistance of Pt100
temperature transmitter (optional): 4 ... 20 mA for temperature measured value, two-wire
Input signalLGC: change in capacitance
Pt100 (optional): change in resistance
temperature transmitter (optional): Pt100 resistance signal, 4-wire
Long-term driftLGC, Pt100 (optional): В± 0.1 % of upper range value (URL) per year
temperature transmitter (optional): В± 0.1 K per year
Drinking water approvalKTW certificate and NSF approval
Degree of protectionLGC, Pt100 (optional): IP68, permanently hermetically sealed, optional terminal box IP66/IP67
temperature transmitter (optional): IP00, moisture condensation permissible, when mounted in the optional terminal boxes IP66/IP67
Influence of medium temperature- thermal change in zero signal and output span for typical temperature range 0 ... 30 В°C (273 ... 303 K):
В± 0.4 % (В± 0.5 %)* of the measuring span
- thermal change in zero signal and output span for the total medium temperature range -10 ... 70 В°C (263 ... 343 K): В± 1.0 % (В± 1.5 %)* of the measuring span (Pt100)
- temperature coefficient (TK) in zero signal and output span: 0.15 %/10 K (0.3 %/10 K)* of the measuring span (Pt100 and temperature transmitter)
*specifications for sensors 0.1 bar and 0.6 bar
Materiallevel probe LGC: 1.4435/316L
process ceramic: Al2O3 aluminum oxide ceramic
seal (internal): EPDM or Viton
protective cap: PE-HD (high-density polyethylene)
terminal box LGC-Z11: PC (polycarbonate)
temperature transmitter LGC-Z13: housing PC (polycarbonate)
extension cable PE: insulation PE (polyethylene), copper wires, twisted
extension cable FEP: insulation FEP (fluorinated ethylene propylene), copper wires, twisted
Construction typedevice with rod probe
Measuring methodLevel control with ceramic measuring cell (dry measuring cell). The pressure acts directly on the rugged ceramic membrane of the LGC level probe and causes it to move by about max. 0.005 mm.
The effects of air pressure on the liquid surface are transferred via a pressure compensating tube through the extension cable to the rear of the ceramic membrane and compensated. Pressure-dependent changes in capacitance caused by membrane movement are measured at the electrodes of the ceramic carrier. The electronics convert the movement into a pressure-proportional signal which is linear to the medium level.
Equipment architectureThe measuring system consists of a LGC level probe and a SMART transmitter power supply
(e. g. KFD2-STC4-Ex1) with a supply voltage between 10 ... 30 V DC.
Dimensionslevel probe LGC: Г (optional): -40 ... 100 В°C (233 ... 373 K)
Reverse polarity protectionReverse voltage protection is integrated in LGC level probe and in the temperature transmitter LGC-Z13 changing the polarities has no impact on operation.
Connection3 terminals in terminal housing as standard
4 terminals in block, accessories LGC-Z15 for conductor cross section 0.08 ... 2.5 mm2

Comments

No comments yet

Write a comment