Pepperl+Fuchs FB5201B

RTD Converter
Pepperl+Fuchs FB5201B
520.03
One click request
ATEX approvalPTB 97 ATEX 1075 ; PTB 97 ATEX 1074 U
Relative humidity95 % non-condensing
Electromagnetic compatibilityNE 21
Shock resistanceshock type I, shock duration 11 ms, shock amplitude 15 g, number of shocks 18
Connectionbackplane bus
Massapprox. 350 g
Vibration resistancefrequency range 10 ... 150 Hz; transition frequency: 57.56 Hz, amplitude/acceleration В± 0.075 mm/1 g; 10 cycles
frequency range 5 ... 100 Hz; transition frequency: 13.2 Hz amplitude/acceleration В± 1 mm/0.7 g; 90 minutes at each resonance
EU-Type Examination CertificatePTB 97 ATEX 1074 U
Storage temperature-25 ... 85 В°C (-13 ... 185 В°F)
Connectionbackplane bus
Measurement range10 ... 400 andOmega; (500 andOmega; incl. line resistance)
Environmental testEN 60068-2-14
Supplementary informationEC-Type Examination Certificate, Statement of Conformity, Declaration of Conformity, Attestation of Conformity and instructions have to be observed where applicable. For information see www.pepperl-fuchs.com.
Shock resistanceEN 60068-2-27
Relative humidityEN 60068-2-56
Degree of protectionIP20 (module) , a separate housing is required acc. to the system description
Damaging gasEN 60068-2-42
Conversion timemax. 20 ms without LFD max. 150 ms with LFD
Degree of protectionIEC 60529
Connectionremovable front connector with screw flange (accessory)
wiring connection via spring terminals (0.14 ... 1.5 mm2) or screw terminals (0.08 ... 1.5 mm2)
Rated voltage12 V DC , only in connection with the power supplies FB92**
Power dissipation0.4 W
Slide-wire sensor10 ... 400 andOmega;
Linearity error0.1 %
Connection2-wire connection: 5, 6
3-wire connection: 1, 5, 6
4-wire connection: 1, 2, 5, 6
Smallest span20 andOmega; for 0.1 % accuracy
LED indicationLED green: supply
LED red: line fault
EAC approvalRussia: RU C-IT.MIII06.B.00129
Damaging gasdesigned for operation in environmental conditions acc. to ISA-S71.04-1985, severity level G3
Line fault detectioncan be switched on/off for each channel via configuration tool
Ambient temperature-20 ... 60 В°C (-4 ... 140 В°F)
Power consumption0.4 W
System informationThe module has to be mounted in appropriate backplanes and housings (FB92**) in Zone 1, 2, 21, 22 or outside hazardous areas (gas or dust). Here, observe the corresponding EC-type examination certificate.
Occupied slots1
Interfacemanufacturer-specific bus to standard com unit
Vibration resistanceEN 60068-2-6
Lead resistancemax. 50 andOmega; per strand
Number of channels1
Codingoptional mechanical coding via front socket
Dimensions28 x 107 x 132 mm (1.1 x 4.2 x 5.2 inch)
Measuring current200 andmicro;A