Endress+Hauser Liquid analysis
20 analog (max) (U, I, TC, RTD, frequency, pulse)
20 HART (max) (HART, 4...20mA)
14 digital (max) (control, event and counting inputs)
12 relays (max)
2 analog (max) (I, pulse)
1 transmitter power supply
TFT 7" colour graphic display
TFT 7" colour graphic display with touch panel
internal memory
SD card
USB flash drive
Mass/heat quantity
heat quantity difference
0.1 to 10 mg/l total chlorine
Process temperature5 to 45 В°C
(41 to 113 В°F)
1 to 4 bar (14.5 to 58 psi)
Measuring methodTotal chlorine consists of free chlorine (HOCl, OCl-) and bound chlorine (chloramines)
All components are reduced at the working electrode
Almost pH independent
Trace: 0 to 5 mg/l ClO2
Standard: 0 to 20 mg/l ClO2
High: 0 to 200 mg/l ClO2
0 to 55 В°C , non-freezing
(32 to 130 В°F)
Max. 2 bar abs
(Max. 29 psi abs)
Trace: 0 to 5 mg/l HOCl
Standard: 0 to 20 mg/l HOCl
High: 0 to 200 mg/l HOCl
0 to 55 В°C (32 to 130 В°F), non-freezing
Process pressureMax. 1 bar (max. 14.5 psi)
Measuring methodClosed, membrane covered measuring cell
Reduction of free chlorine at the cathode
Trace: 0 ... 5 mg/l HOBr
Standard: 0 ... 20 mg/l HOBr
High: 0 ... 200 mg/l HOBr
+0...55 В°C (32 ... 130В°F), non-freezing
Process pressureMax. 1 bar (max. 14.5 psi)
0 to 2 mg/l
Process temperature+0 to 45 В°C (32 to 110 В°F), non freezing
Process pressure1 bar (14.5 psi), 2 bar (29 psi) absolute
Measuring method- closed 2 electrode system
- conversion of ozone into a signal current in nA
- signal current is proportional to the concentration of ozone
- the measurement is not pH dependent
5 ОјS/cm to 200 mS/cm
Process temperature0 to 100 В°C (32 to 212 В°F)
Process pressure1 bar, not intended for continues measurment in process
k=0,01: 0.04 to 20 ВµS/cm
k=0,1: 0.10 to 200 ВµS/cm
Threaded with fixed cable:
-20 to 100 В°C (-4 to 212 В°F)
Threaded with plug-in head:
-20 to 120 В°C (-4 to 248 В°F)
Sterilization: max. 140 В°C (284 В°F) for 30 minutes
13 bar at 20 В°C (188 psi at 68 В°F) absolute
1 bar at 120 В°C (14 psi at 248 В°F) absolute
k=0,1: 0,04 to 500 ВµS/cm
Process temperature-5 to 120 В°C (23 to 248 В°F)
For Sterilizsation: max. 150 В°C at 5 bar
(Max. 302 В°F at 73 psi)
13 bar at 20 В°C (188 psi at 68 В°F)
9 bar at 120 В°C (130 psi at 248 В°F)
k=1: 10 ВµS/cm to 20 mS/cm
Process temperature-20 to 135 В°C at 3.5 bar abs
(-4 to 275 В°F at 50 psi)
17 bar abs at 20 В°C
(246 psi at 68 В°F)
1 ВµS/cm to 500 mS/cm
Process temperature-5 to 120 В°C (23 to 248 В°F)
Sterilization: max. 140 В°C at 6 bar for max. 45 min
(Max. 284 В°F at 87 psi for max. 45 min)
17 bar abs at 20 В°C (247 psi at 68 В°F)
9 bar abs at 120 В°C (131 psi at 248 В°F)
1 ВµS/cm to 500 mS/cm
Process temperature-5 to 120 В°C (23 to 248 В°F)
Sterilization: max. 140 В°C at 6 bar for max. 45 min
(Max. 284 В°F at 87 psi for max. 45 min)
17 bar abs at 20 В°C (247 psi at 68 В°F)
9 bar abs at 120 В°C (131 psi at 248 В°F)
0 to 200 %SAT
Process temperature5 to +60 ЛљC (23 to 140 ЛљF)
Process pressure1 bar, not intended for continues measurment in process
Standard sensor: 0 to 60 mg/l, 0 to 600 %SAT, 0 to 1200 hPa, 0 to 100 Vol%
Trace sensor: 0 to 10 mg/l, 0 to 120 %SAT, 0 to 250 hPa, 0 to 25 Vol%
-5 to 135 В°C (23 to 275 В°F)
Process pressure1 to 12 bar abs (14.5 to 174 psi)
0.01 to 100 mg/l
0.00 to 1000 %SAT
0 to 2000 hPa
-5 to 60 В°C
(20 to 140 В°F)
Max. 5 bar abs
(Max. 72.5 psi abs)
0.004 to 30 mg/l
0.05 to 330 %SAT
0.1 to 700 hPa
0 to 140 В°C
(32 to 284 В°F)
0.02 to 13 bar abs
(0 to 190 psi)
0.004 to 30 mg/l
0.05 to 330 %SAT
0.1 to 700 hPa
1 to 140 В°C
(32 to 284 В°F)
0.02 to 13 bar abs
(0 to 190 psi)
pH 0 to 14
Process temperatureVersion LH: 0 to 110 В°C (32 to 230 В°F)
Version NN: 0 to 80 В°C (32 to 170 В°F)
1 to 10 bar abs at 80 В°C
(15 to 145 psi at 176 В°F)
-1500mV to +1500mV
Process temperature0 to 80 В°C (32 to 170 В°F)
Process pressure1 to 10 bar abs at 80 В°C
(15 to 145 psi abs at 176 В°F)
pH 0 to 14 (1 to 12 application range)
Process temperature-5 to 80 В°C (23 to 176 В°F)
Process pressure1 bar, not intended for continues measurment in process
pH 0 to 14 (1 to 12 application range)
Process temperature5 to 100 В°C (23 to 212 В°F) (0 to 80 В°C (32 to 176 В°F) application range)
Process pressure1 bar, not intended for continues measurment in process
pH 0 to 14 (1 to 12 application range)
Process temperature5 to 100 В°C (23 to 212 В°F) (0 to 80 В°C (32 to 176 В°F) application range)
Process pressure1 bar, not intended for continues measurment in process
pH 0 to 14
Process temperature0 to 135 В°C (32 to 275 В°F) (0 to 100 В°C (32 to 212 В°F) application range)
Process pressure1 bar, not intended for continues measurement in process
Application A
pH: 1 to 12
Application B
pH: 0 to 14
Application F
pH: 0 to 10
Application A: 15 to 80 В°C (5 to 176 В°F)
Application B: 0 to 135 В°C (32 to 275 В°F)
Application F: 0 to 70 В°C (32 to 158 В°F)
Applications A and B:
0.8 to 17 bar (11.6 to 246.5 psi) absolute
Application F:
0.8 to 7 bar (11.6 to 101.5 psi) absolute
-1500 mV to +1500 mV
Process temperature-15 to 135 В°C
(5 to 275 В°F)
0.8 to 17 bar (11.6 to 246.5 psi) absolute
ORP: 1 500 to 1 500 mV
Application A
pH: 1 to 12
Application B
pH: 0 to 14
Application A: 15 to 80 В°C (5 to 176 В°F)
Application B: 0 to 135 В°C (32 to 275 В°F)
0.8 to 17 bar (11.6 to 246,5 psi) absolute